phone Phone
+622122685913
close
menu

PSC

sql
TL23 Benchtop Turbidity Meter Hach
TL23 Benchtop Turbidity Meter Hach
TL23 Benchtop Turbidity Meter Hach
TL23 Benchtop Turbidity Meter Hach
Hover to Zoom

Jual TL23 Benchtop Turbidity Meter Hach



Spesifikasi TL23 Benchtop Turbidity Meter Hach

TL2300 Tungsten Lamp Turbidimeter Hach

Specifications

Accuracy

Ratio on : ±2 % of reading plus 0.01 NTU from 0 - 1000 NTU, ±5 % of reading from 1000 - 4000 NTU based on formazin primary standard

Ratio off : ±2 % of reading plus 0.01 NTU from 0 - 40 NTU

Air purge

Dry nitrogen or instrument grade air ( ANSI MC 11.1, 1975 )

0.05 L/s at 69 kPa ( 10 psig ); 138 kPa ( 20 psig ) max

Hose barb connection for ⅛-inch tubing

Certifications

CE, KC, RCM

Communication

USB

Data logging

2000 total logs, includes reading log, verification log and calibration log

Dimensions ( H x W x D )

153 mm x 395 mm x 305 mm

Display

17.8 mm colour touch screen

Interface

2 USB-A ports for USB flash drive, external printer, keyboard and barcode scanner

Light source

Tungsten filament lamp

Measurement method

Nephelometric

Measuring range

NTU ( Ratio on ) : 0 - 4000

NTU ( Ratio off ) : 0 - 40

EBC ( Ratio on ) : 0 - 980

EBC ( Ratio off ) : 0 - 9.8

Operating temperature range

0 - 40 °C

Power requirements ( Amps )

3.4 A

Power requirements ( Hz )

50/60 Hz

Power requirements ( Voltage )

100 - 240 V AC

Reading modes

Single, continuous, Rapidly Settling Turbidity, signal averaging on or off, ratio on/off

Regulatory

Meets EPA Method 180.1

Repeatability

±1% of reading or 0.01 NTU, whichever is greater ( under reference conditions )

Response time:Signal averaging off

6.8 seconds / Signal averaging on: 14 seconds ( when 10 measurements are used to calculate the average )

Sample cell compatibility

Round cells 95 x 25 mm ( 3.74 x 1 in. ) borosilicate glass with rubber-lined screw caps

Note

Smaller sample cells ( less than 25 mm ) can be used when a cell adapter is used.

Sample requirements

25 mm sample cell : 20 mL minimum

0 to 70 °C ( 32 to 158 °F )

Stabilisation time:Ratio on

30 minutes after start-up

Ratio off

60 minutes after start-up

Storage conditions

-20 - 60 °C

Units

NTU and EBC

Weight

3.0 kg

Alat Ukur Kekeruhan